PerkinElmer Optima 8300 ICP-OES
PerkinElmer Optima 8300 ICP-OES
Key facts
Status
Available
Location
YE403
Room
YE403
Is resource specific training required?
Yes
Minimum reservation length
1 h
Maximum reservation length
8 h
Reservation change/cancellation limit
24 h
Administrator contact info
Name
Email
Phone
Antti Tiihonen
antti.j.tiihonen@jyu.fi
+358505213380
Ari Väisänen
ari.o.vaisanen@jyu.fi
+358408053719
Siiri Perämäki
siiri.e.peramaki@jyu.fi
+358408053495
Kaisa Lampinen
kaisa.m.lampinen@jyu.fi
+358406528384
Pricing
JYU Chemistry and related groups: free of charge
JYU Internal: collaboration required
Commercial use: contact ari.o.vaisanen@jyu.fi or kaisa.m.lampinen@jyu.fi for a quotation
Resource information
Manufacturer
PerkinElmer
Model
Optima 8300
Description
Moodle course
Properties
Instrument installation requirements:
Physical size (instrument) WxDxH 150x80x76 cm, 185 kg
Floor space (instrument and autosampler) WxD 300x125 cm
Physical size (chiller) WxDxH 40x70x60 cm, 90 kg
Floor space (chiller) WxD 70x100 cm
Room temperature 20 ± 2 °C, humidity 20–80 %
Heat dissipation 3 kW instrument + 3 kW chiller
Exhaust fume temperature up to 200 °C
Fume extraction capacity min. 5,7 m
3
/min, funnel type duct
Instrument argon, quality ≥ 99,996 % / 4.6
Instrument nitrogen, quality ≥ 99,999 % / 5.0
Liquid argon is recommended
Suitable optics purge gases are argon or nitrogen (N
2
is cheaper), required flow rate 1,0 L/min (low purge, N
2
) and 5.0 L/min (high purge, N
2
) or argon respectively 1,4 L/min (low purge) and 7,0 L/min (high purge) @ 3,5 bar
Available argon pressure at the regulator should be between 5,5 - 8,3 bar, connector ¼" Swagelok
Argon consumption while plasma on is approximately 10 L/min and in standby/idle 2 L/min, therefore consumption in a week is approximately 40 m
3
if the instrument is running 8 h each workday and is on standby during nights and weekends (shutdown during weekend saves approx. 6 m
3
), and 160 L argon tank lasts for 3-4 weeks
If nitrogen is used, regulator pressure should be between 2,8 - 8,3 bar
Pressurized air (cooling and shear gas) should be clean, dry and oil-free with consumption of 25 L/min @ min. 5,5 bar, ¼" Swagelok (also nitrogen with same requirements can be used)
Electrical connection requirement for the instrument (2,5 m cord, IEC 309 16 A plug "blue, three prongs"): single phase 200-254 V, 50/60 Hz (± 1 %), dedicated 16 A fuse, power 2800 VA
Electrical requirements for the chiller unit (2 m cord, Schuko): single phase 230-240 V, 11 A, 50 Hz, dedicated 16 A fuse, cooling capacity 2850 W @ 20 °C, water circulation 4 gal/min @ 55 psi, requires Polyscience Whispercool heat transfer fluid
GFCI device should not be installed (high earth leakage current)
Computed desk, computer, monitor and their regular electrical connectivity (normal IT equipment)
Communications via USB and Ethernet, no internet connection needed, Windows 10
Image(s)
Host unit
Kemian laitos
Contact information
Name
Email
Phone
Siiri Perämäki
siiri.e.peramaki@jyu.fi
+358408053495
Kaisa Lampinen
kaisa.m.lampinen@jyu.fi
+358406528384
Antti Tiihonen
antti.j.tiihonen@jyu.fi
+358505213380
Ari Väisänen
ari.o.vaisanen@jyu.fi
+358408053719
Capacity
1
Terms of use
Availability
The instrument is available according to the booking calendar.
Do reservations require approvals from administrators?
No
Other terms of use
The prerequisites of the instruments including courses and trainings must be passed before being allowed to conduct analyses independently. See the course link above in resource information section.
Organization
MOTD
Laite on varattavissa kalenterin mukaisesti, mutta tavanomaista vähemmällä käytöllä. Laite voi olla kokonaan pois päältä, joten ota tämä huomioon varatessa (tarkista tilanne). Käytä AVIO-laitetta, jos mahdollista.
The instrument can be reserved according to the calendar, but has less frequent use. It can be in shut down status, so take this into account when making a reservation (check first). Use AVIO, if possible.
Concise description
Inductively coupled plasma - optical emission spectrometer PerkinElmer Optima 8300
×
Research Infrastructure Booking System
Please select an infrastructure:
Front page
Department of Biological and Environmental Science
Digital Services
Department of Physics
Department of History and Ethnology
School of Business and Economics
Department of Chemistry
Department of Language and Communication Studies
Laboratory Center
Faculty of Sport and Health Sciences
Department of Music, Art and Culture Studies
Nanoscience Center
Department of Psychology
Ruusupuisto
Centre for Applied Language Studies
Department of Social Sciences and Philosophy