Beckman Coulter Optima L-90K (2007/2007)
Centrifuge.
Approximate acceleration time: Selection from two profiles: maximum acceleration from 0 rpm to set speed, or slow acceleration from 0 to 500 rpm, followed by maximum acceleration to set speed.
Approximate deceleration time: Select from three profiles: full dynamic braking to 0 rpm, slow (full brake to 500 rpm followed by reduced braking to a gentle stop), or no brake.
Noise Level 1 meter in front of centrifuge: 57 dBA
Set speed: Actual rotor speed ± 20 rpm of the set speed (above 1000 rpm)
Speed range: 1000 rpm to maximum speed
Set time: Up to 99 hr and 59 min; HOLD for runs of unspecified length.
Set temperature: 0 to 40°C in increments of 1°C
Temperature control: ± 0.5°C of set temperature; within 0.3°C of set after equilibration
Operating modes: Standard, programmed, or delayed start; instrument memory can store up to nine programs. Key switch selects NORMAL or ZONAL operation.
Vacuum: Diffusion pump in series with a mechanical pump typically reduces chamber pressure to below 0.7 Pa.